Model Number 24502082
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/31/2018
Event Type  malfunction  
Manufacturer Narrative
The returned product was tested per ameda engineering protocol to determine whether the allegation of leaking fluid could be confirmed.The returned product was assessed for indications of malfunction or thermal event.The returned product was assessed for functionality and met functional specifications.Dark grey substance, e.G battery acid, was observed inside the battery compartment.Internally, no signs of foreign substance, burning, melting or charring were observed.The returned batteries were observably damaged.The battery fluid crust was observed on returned batteries.Additional testing confirmed that batteries may leak when the up battery is placed incorrectly, with the positive and negative end reversed.
 
Event Description
Customer contacted ameda, inc.On (b)(6) 2018 to report pumping on the morning of (b)(6) 2018 using battery power.She explained this was the first time using batteries inside the battery compartment.The ac adapter was not plugged into the pump base at the time.Customer states the pump began to slow down so she replaced the 6 aa batteries with new batteries.Customer noted dark fluid leaking from the battery compartment of the pump base after using the new batteries.She did not come in contact with the fluid and reports no injury or burn.Customer was overnight shipped a replacement purely yours breast pump base.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PURELY YOURS
Type of Device
ELECTRIC BREAST PUMP
Manufacturer (Section D)
AMEDA, INC.
485 half day rd.
suite 320
buffalo grove IL 60089
Manufacturer Contact
linda zager
485 half day rd.
suite 320
buffalo grove, IL 60089
8479642620
MDR Report Key7304757
MDR Text Key101576514
Report Number3009974348-2018-00309
Device Sequence Number1
Product Code HGX
UDI-Device Identifier00810725021477
UDI-Public(01)00810725021477(21)
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K973501
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Remedial Action Replace
Type of Report Initial
Report Date 01/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number24502082
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/31/2018
Initial Date FDA Received02/28/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/24/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age36 YR
-
-