| Catalog Number |
H787107323015 |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
02/07/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The reported defective device has been returned to the manufacturer and an investigation into the root cause for event is currently in progress.The results of the device evaluation will be sent via a follow up medwatch.(b)(4).
|
| |
|
Event or Problem Description
|
|
As reported to angiodynamics on (b)(6) 2018, on the (b)(6)2018, the surgeon was performing an angiographic procedure with a soft-vu omni flush when an incident occurred with the catheter in-vivo.The surgeon reported: "omni flush catheter over supracore wire, up & over aortic bifurcation.On retrieving catheter over the wire, was stuck & broke => one piece being the hub, then the shaft, and leaving the shepard's hook' in the patients superior femoral artery".The catheter tip, "shepard's hook" is embedded in the plaque, located in the patient's superior femoral artery.It was reported the defective disposable device is available for return to the manufacturer for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Returned was a sv of 4fx65cm.035 nb pg catheter without the tip (the associate guidewire was not returned for evaluation).A visual examination of the returned device noter the catheter was fractured at 7.5cm from proximal end.It was also noted the catheter shaft and the strain relief were separated.The catheter shaft was kinked and stretched.The device met all manufacturing dimensional specifications for this product.The customer's reported complaint description "catheter broke at the hub and the shaft" has been confirmed.A definitive root cause for the complaint description could not be determined.The description of the failure is consistent with the catheter becoming caught on the guidewire due to a kink in one of the components and excessive stress placed on the catheter to remove it.A review of the device history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.The instructions for use, which is supplied to the end user with states; "never advance or retract an angiographic catheter or guidewire against resistance.This may result in damage to the vessel, the product, or both.Do not attempt to hand straighten the tip of any curved tipped catheters which are furnished with a tip straightener.This may result in damage to the product.Tip straighteners are furnished on all catheters intended to be straightened with the aid of a tip straightener.Always use a guidewire to remove the catheter from the vasculature.Failure to do so may result in damage to the vessel, puncture site, product, or all three.The maximum pressure limit of catheters intended for flush angiography is stated on the catheter package.When using a pressure injector, do not exceed the stated maximum pressure.Catheters intended for selective angiography do not have a maximum pressure limit stated on the catheter package.Typical flow rates of up to 10cc per second are stated with pressure generated to achieve these typical flow rates never advance or retract an angiographic catheter or guidewire against resistance.This may result in damage to the vessel, the product, or both.Angiodynamics angiographic catheters are designed for use with specific guidewire diameters.The recommended maximum guidewire diameter is specified on the catheter label." a review of the angiodynamics complaint system noted no trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Complaint # (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|