MAUDE Adverse Event Report: COOK INC MULTI-SIDEPORT DUAL CHECK VALVE CATHETER INFUSION SET; DQO CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC MULTI-SIDEPORT DUAL CHECK VALVE CATHETER INFUSION SET; DQO CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Catalog Number MCIS-5.0-35-130-15.0
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 02/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Per the initial reporter, the device will not be returned.Pma/510(k) number = preamendment (b)(4).This report includes information known at this time. a follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported that during inspection of the multi-sideport dual check valve catheter infusion set, it was observed that there was a hair inside of the primary package.The device was not used and did not come into contact with a patient.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation: a review of the complaint history, device history record, documentation, quality control, and specifications of the device as well as a photographic inspection were conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, the customer returned photographs of the complaint device.The presence of hair-like foreign matter was confirmed via the inspection of the returned photographs.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record shows no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Based on the information provided, no product returned and the results of our investigation, a definitive root cause could not be determined.We will continue to monitor for similar complaints.Per the quality engineering risk assessment no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MULTI-SIDEPORT DUAL CHECK VALVE CATHETER INFUSION SET
Common Device Name
DQO CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7305829
Report Number1820334-2018-00638
Device Sequence Number2379428
Product Code DQO
UDI-Device Identifier00827002064256
UDI-Public(01)00827002064256(17)190923(10)7267356
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Catalogue NumberMCIS-5.0-35-130-15.0
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/26/2018
Supplement Date Received by Manufacturer05/21/2018
Initial Report FDA Received Date03/01/2018
Supplement Report FDA Received Date06/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-