• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MMJ, S.A. DE C.V. MALLINCKRODT; TUBE, TRACHEAL (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MMJ, S.A. DE C.V. MALLINCKRODT; TUBE, TRACHEAL (W/WO CONNECTOR) Back to Search Results
Model Number 18775
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 02/06/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during cuff test, the inflated cuff leaked air.The customer indicated that there was no patient involved in this event.
 
Additional Manufacturer Narrative
(first name, last name, email address), (phone#).Evaluation summary: without the actual complaint sample a full investigation cannot be carried out therefore we are unable to determine the cause for this specific event.Manufacturing controls were found acceptable and effective to detect the reported failure mode.An inflation/deflation was perform and the operator inspects for no leaks and segregates the defective parts.Device history records were reviewed and no discrepancies were found.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MALLINCKRODT
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
MMJ, S.A. DE C.V.
ave. henequen no 1181 desarrol
ciudad juarez,ci 32590
MX  32590
MDR Report Key7305971
Report Number2936999-2018-00124
Device Sequence Number957438
Product Code BTR
UDI-Device Identifier60884522007608
UDI-Public60884522007608
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K090352
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 08/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/06/2022
Device Model Number18775
Device Catalogue Number18775
Device Lot Number17C0328JZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/23/2018
Initial Date Received by Manufacturer 02/13/2018
Supplement Date Received by Manufacturer08/01/2018
Initial Report FDA Received Date03/01/2018
Supplement Report FDA Received Date08/16/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-