| Model Number |
M00516710 |
| Medical Device Problem Code |
Positioning Problem (3009)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/15/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Patient's exact age is unknown; however it was reported that the patient was over the age of 18.(b)(4).The device has been received for analysis; however the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation on (b)(6) 2018 that a wallflex esophageal fully covered stent was to be used to treat a malignant stenosis in the cardia of the esophagus during a stent placement procedure performed on (b)(6) 2018.Reportedly, the patient¿s anatomy was not tortuous and was not dilated prior to stent placement.According to the complainant, during the procedure, the stent was able to be deployed.However, when the stent was viewed endoscopically to check for placement, the stent was noted to have moved towards the stomach due to peristalsis.The stent was removed with forceps and the procedure was completed with a different device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
| |
|
Additional Manufacturer Narrative
|
|
A deployed wallflex esophageal fully covered stent and delivery system were received for analysis.Visual examination of the returned device identified no issue with the stent and delivery system.The stent was measured to be within specifications.Taking all available information into consideration, the investigation concluded that the reported failure was likely due to anatomical or procedural factors encountered during the procedure which limited the performance of the device.Therefore, the most probable root cause for this complaint is operational context.A labeling review was performed, and from the information available, this device was used per the directions for use (dfu) / product label.A review of the device history record (dhr) confirmed that the device met all material, assembly and product specifications at the time of release to distribution.A search of the complaint database revealed that no other similar complaints exist for the specified batch.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation on (b)(6) 2018 that a wallflex esophageal fully covered stent was to be used to treat a malignant stenosis in the cardia of the esophagus during a stent placement procedure performed on (b)(6) 2018.Reportedly, the patient¿s anatomy was not tortuous and was not dilated prior to stent placement.According to the complainant, during the procedure, the stent was able to be deployed.However, when the stent was viewed endoscopically to check for placement, the stent was noted to have moved towards the stomach due to peristalsis.The stent was removed with forceps and the procedure was completed with a different device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.Corrected information received on april 09, 2018.According to the complainant, the correct procedure date is (b)(6) 2017.
|
| |
|