MEDACTA INTERNATIONAL SA AMISTEM H HA COATED STD STEM SIZE6; CEMENTLESS HIP STEM Back to Search Results
Catalog Number 01.18.136
Health Effect - Clinical Code Joint Disorder (2373)
Date of Event 01/30/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Clinical evaluation performed by medical affairs on 16 february 2018: hip revision surgery occurred 5,5 years after primary cementless total hip arthroplasty.Radiographic image provided shows the presence of a mobilized stem.Aseptic loosening is a possible, literature described adverse event after cementless tha and causes are often unknown.In this case, reason of failure cannot be determined.Batch review performed on 26 february 2018 lot 120924: (b)(4) items manufactured and released on 22 june 2012.Expiration date: 2017-05-31 no anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.
 
Event or Problem Description
The patient came in complaining of instability 5 years and 5 months after primary.The cause is unknown.The stem came out very easily, took no time with the medacta stem extractor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMISTEM H HA COATED STD STEM SIZE6
Common Device Name
CEMENTLESS HIP STEM
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ  6874
MDR Report Key7306125
Report Number3005180920-2018-00102
Device Sequence Number1261015
Product Code LZO
UDI-Device Identifier07630030804120
UDI-Public07630030804120
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K093944
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2017
Device Catalogue Number01.18.136
Device Lot Number120924
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/30/2018
Initial Report FDA Received Date03/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age49 YR
-
-