• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SI-BONE, INC. IFUSE IMPLANT SYSTEM; ORTHOPEDIC ROD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SI-BONE, INC. IFUSE IMPLANT SYSTEM; ORTHOPEDIC ROD Back to Search Results
Model Number 7055-90
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code Undesired Nerve Stimulation (1980)
Date of Event 02/08/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Based on the information provided, review of the surgical technique manual, ifu, certificates of conformance and fmea, there is no indication of device failure and no indication that the device was out of specification.The most probable root cause is improper implant size selection.
 
Event or Problem Description
The patient had bilateral si joint arthrodesis in (b)(6) 2015 where three implants were placed on each side.The patient reported leg pain complaints to the surgeon after the initial procedure.The patient later presented to a different surgeon who determined that most superior positioned left side implant had breached the neuroforamen.In (b)(6) 2018, the surgeon performed a revision procedure where he removed the implant using chisels as it was solidly fixed in bone with heavy bone growth on the implant.He then replaced it with a shorter ifuse implant.The patient's pain symptoms resolved following the revision procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IFUSE IMPLANT SYSTEM
Common Device Name
ORTHOPEDIC ROD
Manufacturer (Section D)
SI-BONE, INC.
3055 olin avenue
suite 2200
san jose CA 95128 2066
MDR Report Key7306669
Report Number3007700286-2018-00018
Device Sequence Number2300384
Product Code OUR
UDI-Device Identifier00859256003316
UDI-Public00859256003316
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K080398
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/20/2020
Device Model Number7055-90
Device Lot Number373334
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/08/2018
Initial Report FDA Received Date03/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
-
-