| Model Number |
M00562531 |
| Medical Device Problem Code |
Failure to Cut (2587)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.However, the complainant reported that the device was not expired.(b)(4).Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a profile extra small oval snare was used during a cold polypectomy procedure.According to the complainant, during the procedure, the snare loop failed to cut the target polyp.The procedure was completed with a different device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be good.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results: visual evaluation of the returned device revealed that the catheter was twisted and torn near the handle section.Functional analysis showed that the device failed to extend due to the condition of the returned device.The reported complaint that the snare loop failed to cut was confirmed.Due to anatomical/procedural factors encountered during the procedure; performance of the device was limited.Therefore, the most probable root cause classification is operational context.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a profile extra small oval snare was used during a cold polypectomy procedure.According to the complainant, during the procedure, the snare loop failed to cut the target polyp.The procedure was completed with a different device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be good.
|
| |
|