Model Number 7035-90
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code Arthralgia (2355)
Date of Event 02/14/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Based on the information provided, review of the surgical technique manual, ifu, certificates of analysis and fmeas, there is no indication of device failure and no indication that the device was out of specification.The most probable root cause is a malpositioned implant.
 
Event or Problem Description
The patient had right side si joint arthrodesis in (b)(6) 2016 where three implants were placed.The patient later complained of continued si joint pain.The surgeon determined that the most caudal positioned implant was malpositioned and showed lucency.In (b)(6) 2018, the surgeon performed a revision surgery where he removed the implant and placed a longer ifuse implant in a more ventral position.The status of the patient following the revision is not known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IFUSE IMPLANT SYSTEM
Common Device Name
ORTHOPEDIC ROD
Manufacturer (Section D)
SI-BONE, INC.
3055 olin avenue
suite 2200
san jose CA 95128 2066
MDR Report Key7306790
Report Number3007700286-2018-00011
Device Sequence Number1066401
Product Code OUR
UDI-Device Identifier00859256003279
UDI-Public00859256003279
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K080398
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/08/2020
Device Model Number7035-90
Device Lot Number493557
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/14/2018
Initial Report FDA Received Date03/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
-
-