| Model Number |
42648-06 |
| Medical Device Problem Code |
Detachment Of Device Component (1104)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
One (1) used transpac iv monitoring kit was received and evaluated.A tubing separation was observed in the 3" arterial pressure tubing and the safeset port.The cause of the separation was insufficient solvent bond on the tubing.
|
| |
|
Event or Problem Description
|
|
The customer reported the arterial line became disconnected at a junction point during patient use.It was reported that the monitor was able to sense an arterial line disconnection.There was a small amount of blood loss reported but no report of adverse event or delay in critical therapy.
|
| |
|
Additional Manufacturer Narrative
|
|
Subsequent to the initial medwatch, a user facility medwatch was received with additional information.
|
| |
|
Event or Problem Description
|
|
The event occurred during blood pressure monitoring.
|
| |
|