• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - COSTA RICA (COYOL) JAGWIRE¿; ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - COSTA RICA (COYOL) JAGWIRE¿; ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY Back to Search Results
Model Number M00556591
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/14/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device has been received for analysis.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a jagwire¿ guidewire was used in the common bile duct during a lithotripsy procedure performed on (b)(6) 2018.According to the complainant, during the procedure as they were capturing the stones in the left intrahepatic bile duct, cannulation was performed and was passed through the stones.Contrast imaging was then performed.After that, as they tried to exchange it but the guidewire could not be moved and was stuck in the stones.They tried to moved it several times but met with resistance the core wire was broken as well as the hydrophilic tip detached and remained on the distal part of the stones.They replaced it with a different guidewire and attempted to retrieve the stones using a retrieval balloon, however the retrieval balloon would not pass through the stones.The procedure was completed with a different device.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
A visual examination of the revealed that the distal tip was partially detached, exposing the corewire.Presence of adhesive remnants were found indicating that the pebax was properly attached to the corewire.The length of the tip is 5.0cm and does not present broken sections.The complaint is not consistent with the returned device that the core wire was broken.Probably procedural or anatomical factors encountered during the procedure could have affected the device performance and its integrity.It is most likely that during procedure the device could have been manipulated, since the failure found is an issue that could have been generated by excessive manipulation of the device, also interaction with other devices might have impacted the integrity of the device during the procedure.Therefore, the complaint conclusion investigation code is operational context since probably due to anatomical/procedural factors encountered during the procedure, performance was limited.A dhr (device history record) review was performed and no deviation was found.A search of the complaint database confirmed that no similar complaints exist for the specified batch.
 
Event or Problem Description
It was reported to boston scientific corporation that a jagwire¿ guidewire was used in the common bile duct during a lithotripsy procedure performed on (b)(6) 2018.According to the complainant, during the procedure as they were capturing the stones in the left intrahepatic bile duct, cannulation was performed and was passed through the stones.Contrast imaging was then performed.After that, as they tried to exchange it but the guidewire could not be moved and was stuck in the stones.They tried to moved it several times but met with resistance the core wire was broken as well as the hydrophilic tip detached and remained on the distal part of the stones.They replaced it with a different guidewire and attempted to retrieve the stones using a retrieval balloon, however the retrieval balloon would not pass through the stones.The procedure was completed with a different device.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JAGWIRE¿
Common Device Name
ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS 
MDR Report Key7307776
Report Number3005099803-2018-00660
Device Sequence Number11075131
Product Code OCY
UDI-Device Identifier08714729501398
UDI-Public08714729501398
Combination Product (Y/N)N
PMA/510(K) Number
K141820
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/10/2020
Device Model NumberM00556591
Device Catalogue Number5659
Device Lot Number20147235
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/19/2018
Initial Date Received by Manufacturer 02/14/2018
Supplement Date Received by Manufacturer03/10/2018
Initial Report FDA Received Date03/01/2018
Supplement Report FDA Received Date03/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-