| Catalog Number |
RBYPOD5 |
| Medical Device Problem Codes |
Inadequacy of Device Shape and/or Size (1583); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/06/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Results: the pod5 pusher assembly was kinked.Conclusions: evaluation of the returned pod5 revealed that the pusher assembly was kinked.This damage was likely incidental and may have occurred from packaging the device for return to penumbra.Due to the returned condition of the pod5, the device could not be functionally tested.Therefore, the root cause of the reported difficulty advancing could not be determined.The lantern referred to in the complaint was not returned for evaluation.Penumbra coils are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report numbers: 3005168196-2018-00454.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure to treat a large varicocele using ruby coils and a pod5.During the procedure, prior to use, a ruby coil pusher assembly became kinked while being removed from the dispenser hoop; therefore it was set aside and the procedure continued with other coils.The physician then had difficulty advancing a pod5 into the existing coil mass in the varicocele due to the fact that the pod5 was the wrong size; therefore, it was removed.The procedure was successfully completed using other ruby coils and the same lantern delivery microcatheter (lantern).There was no report of an adverse effect to the patient.
|
| |
|
Search Alerts/Recalls
|