Model Number S7 |
Device Problems
No Display/Image (1183); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 02/06/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
No patient information provided as no patient was involved in this concern.A medtronic representative inspected the navigation system onsite and confirmed that there was no signal going to the surgeon monitor.Additionally, the uninterruptible power supply (ups) was not functioning when the system was unplugged.The power supply was replaced and the issue resolved.The representative also replaced the uninterruptible power supply (ups) and ups batteries.A full system check-out was completed following part replacement and all tests passed.Full system functionality was confirmed and the system was returned to service.No parts have returned to the manufacturer for evaluation.
|
|
Event Description
|
A medtronic representative reported that while outside of a procedure, the surgeon monitor on the navigation system was not getting power.The representative tested the output on the multi-out power supply and was not getting any voltage back form the monitor input.There was no patient present when this issue was observed.
|
|
Manufacturer Narrative
|
The uninterruptible power supply (ups) was returned to the manufacturer for evaluation.When attached to a known operational battery, the ups would power down immediately.It was noted that all outputs measure normally once the ac was disconnected.The power supply of the imaging system was returned to the manufacturer for evaluation.Testing found that there was a fuse that was open on the board of the power supply.
|
|