| Model Number |
MICL13.2 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); No Code Available (3191)
|
| Date of Event |
01/15/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Patient code: (b)(4) secondary surgical intervention, lens explanted and exchanged for shorter len- work order search: no similar complaint was reported for units within the same lot.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated the surgeon implanted a 13.2mm micl13.2 implantable collamer lens, -12.5 diopter, in the patient's eye on (b)(6) 2018.The lens was explanted on (b)(6) 2018 due to excessive vaulting and was exchanged for a shorter lens.The lens was discarded.Additional information has been requested but none has been forthcoming.If additional information is received, a supplemental medwatch will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
The lens was explanted on (b)(6) 2018 due to excessive vaulting and blurred vision.The lens was exchanged for a shorter lens and the problem was resolved.Patient's post-op status of the eye appears normal, good outcome and patient is pleased.(b)(4).
|
| |
|
Search Alerts/Recalls
|