• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT OMNILINK ELITE PERIPHERAL STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT OMNILINK ELITE PERIPHERAL STENT SYSTEM Back to Search Results
Catalog Number 1012625-39
Health Effect - Clinical Code Vessel Or Plaque, Device Embedded In (1204)
Date of Event 02/07/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The device was not returned for analysis.Possible factors that may contribute to difficult to position through an introducer sheath may include, but are not limited to: damage to the stent, damaged to the device, damage to the introducer sheath.In this case, it is possible that the outer diameter of the introducer sheath may have been too large that during advancement the introducer sheath encountered resistance and/or became entrapped with the previously implanted stent resulting in the reported stretched stent implant; however, this cannot be confirmed.The investigation was unable to determine a conclusive cause for the reported difficult to position.The reported material deformation and additional therapy non-surgical treatment appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to design, manufacture or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was to treat a moderately calcified lesion in the right common iliac artery.A 9x39mm omni elite stent was deployed in the right common iliac artery; however, a sheath was being advanced through the stent to obtain left access and the sheath became stuck with the stent.The stent migrated and elongated towards the left common iliac artery and aorta bifurcation.The procedure was successfully completed by using two unknown kissing stents to smash the omnilink elite stent against the vessel wall.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNILINK ELITE PERIPHERAL STENT SYSTEM
Common Device Name
PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7308279
Report Number2024168-2018-01502
Device Sequence Number19395493
Product Code NIO
UDI-Device Identifier08717648179341
UDI-Public08717648179341
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P110043
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 03/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2021
Device Catalogue Number1012625-39
Device Lot Number7090441
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/07/2018
Initial Report FDA Received Date03/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-