The device was not returned to olympus for evaluation.The cause of the reported complaint cannot be confirmed.As a preventive measure, the instruction manual for the device has directions for pre-procedure visual and functional inspection, including verification that the needle distal tip extends and retracts into the sheath.The instruction manual also has warnings and cautions for preventing breakage at the proximal end: ¿do not apply bending force to the handle section.¿ ¿if you feel excessive resistance while operating the needle, do not push the needle slider forcibly.¿ and ¿if you feel excessive resistance while operating the needle adjuster lever, release the needle adjuster lever once and fix it again.Otherwise, it could damage the needle adjuster lever¿.The instruction manual also advises, ¿always have a spare instrument available in case the primary instrument malfunctions.¿.
|
This supplemental report is being submitted to provide the device evaluation results, update the lot number.The device was returned to olympus for evaluation.The evaluation confirmed the customer complaint for the needle broken within the device and the needle stuck in position.The evaluation found a sharp kink in the sheath of the insertion portion below the metal protector boot.This sharp kink had caused the inner metal tube to break, resulting in the needle not being able to retract.The probable cause for the damage is external applied stress and force.
|