| Model Number |
27703 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/03/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported the catheter was leaking.No patient symptoms or complications were reported and the patient¿s status was listed as no injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Approximately 33 cm of the edm ventricular catheter was returned.The catheter was patent and met the requirements for leak testing.Therefore the conditions of the complaint could not be duplicated by laboratory personnel.Proteinaceous debris was observed within the interior and exterior of the catheter.All catheters are 100% inspected at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information received reported the cause of the leaking was not determined.
|
| |
|