| Model Number |
ZMB00 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown, not provided, but the best estimate date is between (b)(6) 2015 and (b)(6) 2017.(b)(4).Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a tecnis multifocal intraocular lens (model zmb00 +22.0 diopter) was explanted in a secondary procedure due to the patient experiencing dysphotopsia.A tecnis monofocal intraocular lens (model z9002 +21.0 diopter) was implanted as the replacement lens.The patient was reported to be doing well at discharge.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: device available for evaluation ¿ yes, returned to manufacturer on 3/15/2018.Device returned to manufacturer ¿ yes.Device evaluation: the intraocular lens (iol) was returned to the manufacturing site for evaluation.Visual inspection that was performed by a qualified operator using 12x magnification shows the product is cut in half, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.The customer's reported complaint was not verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|