The hawkone was not received for evaluation.No ancillary devices or cine images from the procedure were received.A single photographic image was received.The photographic image shows the hawkone cutter separated just distal of the distal assembly cutter window.A tearing of the distal assembly just distal of the cutter window was observed.The coil housing, flushing port, and distal tip remain on the guidewire that protrudes from the support catheter.The cutter head and drive shaft extend distally from the cutter window.Due to the quality, clarity, and pixelation of the photographic image it is not possible to confirm or not confirm if the guidewire prolapsed.If information is provided in the future, a supplemental report will be issued.
|