• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN ENDO GIA; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN ENDO GIA; STAPLE, IMPLANTABLE Back to Search Results
Model Number 030450
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 03/17/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This event was previously reported to the fda on a summary report and is now being submitted on a 3500 a per fda request.Evaluation summary: post market vigilance (pmv) led an evaluation of the device.The visual inspection and functional evaluation of the device had acceptable results.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter: the anvil cannot be opened when reloading the handle.The problem was noticed when testing prior to use.There was no patient involvement and no patient injury.The current condition of the patient is stable.No medical intervention was required and the surgery time was not extended by thirty minutes or more.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GIA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key7308854
Report Number1219930-2018-01188
Device Sequence Number16514943
Product Code GDW
UDI-Device Identifier10884523002812
UDI-Public10884523002812
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K111825
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2020
Device Model Number030450
Device Catalogue Number030450
Device Lot NumberN5J1077KX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/17/2017
Initial Report FDA Received Date03/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-