• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL OT VERIO TEST STRIPS; GLUCOSE MONITORING SYS/KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL OT VERIO TEST STRIPS; GLUCOSE MONITORING SYS/KIT Back to Search Results
Lot Number 4268091
Device Problem Sticking (1597)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
See incident description for device evaluation.
 
Event Description
Device evaluation: the lay user/patients test strips have been returned and evaluated by lifescan (lfs) product analysis with the following findings: the test strips failed testing and the reported issue was confirmed.The test strips were found to be stuck together due to drops of adhesive from the slitting and vialling process.On (b)(6) 2017, the lay user/reporter contacted lfs (b)(4), alleging that their onetouch verio test strips were stuck together.This complaint was initially ruled out because there was no indication that the product malfunctioned during customer service troubleshooting and there was also no allegation of an adverse event as a result of the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OT VERIO TEST STRIPS
Type of Device
GLUCOSE MONITORING SYS/KIT
Manufacturer (Section D)
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
gubelstrasse 34
zug 6300
SZ  6300
Manufacturer (Section G)
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
gubelstrasse 34
zug 6300
SZ   6300
Manufacturer Contact
m. chiusano / y. wandolski
gubelstrasse 34
zug 6300
SZ   6300
MDR Report Key7309022
MDR Text Key101610535
Report Number3008382007-2018-00689
Device Sequence Number1
Product Code NBW
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K093745
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other Caregivers
Type of Report Initial
Report Date 02/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Lot Number4268091
Other Device ID Number1-SPA5A5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/01/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date02/07/2018
Date Manufacturer Received02/07/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age63 YR
-
-