MAUDE Adverse Event Report: COOK ENDOSCOPY D.A.S.H. DOMETIP DOUBLE LUMEN SPHINCTEROTOME; KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK ENDOSCOPY D.A.S.H. DOMETIP DOUBLE LUMEN SPHINCTEROTOME; KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Catalog Number DASH-21-480
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/02/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation evaluation: an evaluation of the returned device confirmed the report on wire guide coating damage.The associated sphincterotome was not included in the return.The wire guide has coating damage near the distal end.Coating has peeled exposing bare core wire between 1.8 cm to 4.2 cm from the distal end and the coating has bunched up at 1.5 cm from the distal end.A small portion of the coating was removed at the proximal end of the coating to determine if the coil spring was still present.This review found the coil spring to be present in the intended location, however due to the condition of the coating on the returned device, it cannot be determined if any section of the coating is missing.A discrepancy or anomaly that could have contributed to the reported event was not observed during our laboratory analysis of the returned product.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.The instructions for use advise the user: "for best results, wire guide should be kept wet." the technique described includes flushing the endoscope accessory channel and/or lumen of accessory device with sterile water before wire guide insertion.If this flushing technique is not followed, this can contribute to wire guide coating damage.Prior to distribution, all d.A.S.H.Dometip double lumen sphincterotomes are subjected to a visual inspection and functional test to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event or Problem Description
During an endoscopic retrograde cholangiopancreatography (ercp), the physician used a cook d.A.S.H.Dometip double lumen sphincterotome.The wire coating began to strip off near the distal end.No portion came off in the patient.The procedure was completed with another wire.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
D.A.S.H. DOMETIP DOUBLE LUMEN SPHINCTEROTOME
Common Device Name
KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
MDR Report Key7309097
Report Number1037905-2018-00079
Device Sequence Number877500
Product Code KNS
UDI-Device Identifier00827002257900
UDI-Public(01)00827002257900(17)191026(10)W3782124
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K901443
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberDASH-21-480
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date02/02/2018
Device Age4 MO
Event Location Hospital
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/05/2018
Initial Report FDA Received Date03/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
-
-