• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MALLINCKRODT MEDICAL SHILEY; TUBE, BRONCHIAL (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MALLINCKRODT MEDICAL SHILEY; TUBE, BRONCHIAL (W/WO CONNECTOR) Back to Search Results
Model Number 125037
Medical Device Problem Code Deflation Problem (1149)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 02/06/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during testing, the inflated tracheal cuff did not deflate.The customer indicated that there was no patient involvement.
 
Additional Manufacturer Narrative
Evaluation summary: one sample returned for evaluation contained in its original opened unit pack.Visual examination shows that the shape of the tubing is slightly altered using the stylet wire.The stylet wire was removed from the tubing and air was removed from the tracheal cuff and from the bronchial.The stylet wire was replaced into the tubing.The returned unit was inflated and deflated without any issue.The stylet wire was removed from the tubing and re-inflated without the stylet wire.The returned unit inflated/deflated without any issue.The returned unit inflated/deflated with and without the stylet wire and no restriction was noted.The reported defect could not be detected on the unit returned for evaluation.The most probable root cause is the adaptor was not fully depressed during the deflation stage of the process.All of our products undergo a four hour inflate/deflate test and it is at this stage of the process that any defects are removed from the lot.The reported defect could not be detected on the unit returned for evaluation.Information has been added to the database and trends will continue to be monitored.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE, BRONCHIAL (W/WO CONNECTOR)
Manufacturer (Section D)
MALLINCKRODT MEDICAL
cornamaddy
athlone 3810
MDR Report Key7309423
Report Number8020889-2018-00026
Device Sequence Number16416702
Product Code BTS
UDI-Device Identifier30884522008741
UDI-Public30884522008741
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K771219
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 06/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/01/2022
Device Model Number125037
Device Catalogue Number125037
Device Lot Number201702369X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/29/2018
Initial Date Received by Manufacturer 02/13/2018
Supplement Date Received by Manufacturer05/30/2018
Initial Report FDA Received Date03/02/2018
Supplement Report FDA Received Date06/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-