| Model Number |
XTP114IJSE |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An investigation has been initiated by the contract manufacturer.We are waiting for the sample to be returned for evaluation.
|
| |
|
Event or Problem Description
|
|
Prior to the start of dialysis the dressing was opened and the catheter suture wing was found broken.
|
| |
|
Additional Manufacturer Narrative
|
|
The supplier opened an internal capa to perform an in-depth investigation.The evaluation of the returned samples and retained samples indicated that the material was exposed to excessive heat.Root cause: temperature fluctuation on the boy molding machine contributed to the reported failure.Corrective action plan: the suture wing will no longer manufactured on the boy molding machine.All manufacturing will be performed on the toyo machine.Preventative action plan: update to the inspection and workmanship standard to provide instructions for additional visual and functional inspection methods.
|
| |
|