| Catalog Number |
1C8687 |
| Medical Device Problem Codes |
Degraded (1153); Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that an unspecified number of clearlink blood bag spike adapters were leaking.The adapters leak when they are used to push the reagent, dimethyl sulfoxide (dmso).It was reported that the plastic housing surrounding the rubber port of the set deteriorate/breaks apart after the dmso is pushed through which results in leaks.The reporter advised that they do not use the blood bag spike adapter for iv access within their facility.Their operations staff uses the adapter to make a cryopreservation media for their stem cell products and the dmso protects the cells during the freezing process.So, the staff had been transferring the dmso from its original container to another bag to store the cryoprotectant.There was no patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
The actual device was not available; however, a photograph of the sample was provided for evaluation.Visual inspection of the photograph revealed the plastic housing which surrounds the rubber port of the set was deteriorated.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Search Alerts/Recalls
|