• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG LEVEL ONE THORACIC; PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG LEVEL ONE THORACIC; PLATE Back to Search Results
Model Number 24-015-22-71
Medical Device Problem Codes Break (1069); Use of Device Problem (1670)
Health Effect - Clinical Code Impaired Healing (2378)
Date of Event 01/26/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated report: mdr 9610905-2018-00011.Reference exemption number e2017029.
 
Event or Problem Description
It was reported that the plate broke in a patient with pre-operative bone sarcoma.Review of the ct scans show that the lower rib that was plated was not reduced to bone-to-bone contact.The plate on this same rib was not contoured to align on the rib completely.This resulted in screws not being inserted into bone.The screws that were inserted were placed in the inferior border resulting in sub-optimal fixation.Ifu 90-813-59-10 states: "the surgeon must ensure the fractured segments have bone to bone contact.Any bridging of a continuity defect must be combined with a bone graft.".
 
Additional Manufacturer Narrative
An investigation was performed on the basis of complaint statistics as no device was returned for evaluation.The product device history records were also reviewed based on the lot number provided, and no anomalies were found.The root cause for the failure cannot be determined due to no device being returned.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL ONE THORACIC
Common Device Name
PLATE
Manufacturer (Section D)
KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG
kolbinger strabe 10
muhlheim/donau, 78570
GM  78570
MDR Report Key7309538
Report Number9610905-2018-00013
Device Sequence Number2274173
Product Code HRS
UDI-Device Identifier00888118101327
UDI-Public(01)00888118101327
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K153482
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/05/2018
Report Date (Section F) 02/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Model Number24-015-22-71
Device Lot Number33220175
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Event Location Hospital
Date Report to Manufacturer02/05/2018
Initial Date Received by Manufacturer 02/05/2018
Supplement Date Received by Manufacturer02/05/2018
Initial Report FDA Received Date03/02/2018
Supplement Report FDA Received Date03/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age56 YR
-
-