DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number MT22495-PNK
Device Problem Display or Visual Feedback Problem (1184)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/09/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Dexcom was made aware on (b)(4) 2018 that on (b)(6) 2018 the receiver display was frozen.No additional event or patient information is available.No product or data was provided for evaluation.The complaint confirmation of the frozen display was undetermined.The probable cause could not be determined.
 
Manufacturer Narrative
Com-(b)(4).
 
Event Description
The receiver was returned for evaluation.An external visual inspection was performed and passed.A receiver charge and boot was performed and passed.A receiver download was performed passed.A functional test was performed and passed.An interior visual inspection was performed and passed.Confirmation of the complaint was not confirmed.The probable cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence dr.
san diego, CA 92121
8582000200
MDR Report Key7309760
MDR Text Key101522008
Report Number3004753838-2018-021536
Device Sequence Number1
Product Code MDS
UDI-Device Identifier10386270000160
UDI-Public10386270000160
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMT22495-PNK
Device Catalogue NumberSTK-DR-PNK
Device Lot Number5229902
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/05/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received03/07/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/14/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age14 YR
-
-