• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD SOFSILK; SUTURE, NONABSORBABLE, SILK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD SOFSILK; SUTURE, NONABSORBABLE, SILK Back to Search Results
Model Number SS-695
Medical Device Problem Codes Break (1069); Detachment Of Device Component (1104)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 01/22/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter: the issue occurred during open (left) neck dissection and free flap.The tip broke off of the needle -- unrecovered form the field.Suspected to be in the patient, near the mandible.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection and functional evaluation of the sample had acceptable results.A bend moment test was performed on the undamaged needles, and the results exceeded the specifications for this product.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all manufacturers¿ quality release specifications at the time of manufacture.Analysis concluded there were no assembly or component related failures.A review of the device history record indicates the product was released meeting all medtronic quality release specifications at the time of manufacture.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOFSILK
Common Device Name
SUTURE, NONABSORBABLE, SILK
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key7309804
Report Number9612501-2018-00287
Device Sequence Number1276653
Product Code GAP
UDI-Device Identifier20884521088310
UDI-Public20884521088310
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K981128
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 05/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2022
Device Model NumberSS-695
Device Catalogue NumberSS-695
Device Lot NumberD7E0725FX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2018
Initial Date Received by Manufacturer 02/01/2018
Supplement Date Received by Manufacturer04/06/2018
04/25/2018
04/25/2018
Initial Report FDA Received Date03/02/2018
Supplement Report FDA Received Date04/19/2018
05/02/2018
05/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-