Model Number IMMULITE 2000 ANTI-TG AB
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/31/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer does not have any sample left for in-house testing.The cause of the discordant, falsely low anti-tg ab result on one patient sample is unknown.Siemens is investigating the issue.Mdr 2432235-2018-00093 was filed for the same issue.
 
Event or Problem Description
A discordant, falsely low result was obtained for autoantibodies to thyroglobulin (anti-tg ab) assay on one patient sample on an immulite 2000 xpi instrument, while using kit lot 675.The result from the immulite 2000 xpi instrument was reported to the physician(s), who questioned it.The sample was repeated on an alternate platform, resulting higher.It is unknown if the result from the alternate platform was reported to the physician(s).There are no reports of patient intervention or adverse health consequence due to the discordant, falsely low anti-tg ab result.
 
Additional Manufacturer Narrative
The initial mdr 2432235-2018-00092 was filed on 02-mar-2018.Additional information (05-apr-2018): a siemens headquarters support center (hsc) specialist reviewed the event data.The hsc specialist stated that without having a sample to perform in-house testing, knowing which alternate platform was used to perform repeat testing for sample in question or having other thyroid test results, the cause of the issue could not be determined.The potential cause of the issue was due to sample handling/sample preparation.Quality controls and adjustments were valid and with acceptable range.The device is performing within manufacturing specifications.No further evaluation of device is required.Result and conclusion codes have been updated in section h6.Mdr 2432235-2018-00093_s1 was filed for the same issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMMULITE 2000 ANTI-TG AB
Common Device Name
IMMULITE 2000 ANTI-TG AB
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS, LIMITED
glyn rhonwy
llanberis
gwynedd, LL55 4EL
UK  LL55 4EL
MDR Report Key7310279
Report Number2432235-2018-00092
Device Sequence Number1343179
Product Code JZO
UDI-Device Identifier00630414962108
UDI-Public00630414962108
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K991094
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 04/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIMMULITE 2000 ANTI-TG AB
Device Lot Number675
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 02/07/2018
Supplement Date Received by Manufacturer04/05/2018
Initial Report FDA Received Date03/02/2018
Supplement Report FDA Received Date04/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-