Model Number 7205788
Device Problems Circuit Failure (1089); Power Problem (3010)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/01/2018
Event Type  malfunction  
Event Description
It was reported that a short was found with the power cable.There was not delay or patient injury reported.
 
Manufacturer Narrative
The device was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and it was observed that several handpiece connector pins and sockets were corroded.The unit was tested for proper operation and performed as expected.The complaint was not verified and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYONICS POWER CABLE ASSEMBLY
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
MDR Report Key7310499
MDR Text Key101479161
Report Number1643264-2018-00190
Device Sequence Number1
Product Code HAB
UDI-Device Identifier03596010408990
UDI-Public(01)03596010408990
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7205788
Device Catalogue Number7205788
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received04/03/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse