| Model Number |
8140-12-0.8 |
| Medical Device Problem Codes |
Partial Blockage (1065); Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/18/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product involved in the report has been returned and is being processed for evaluation.A review of the device history record is not possible as no lot number was provided.Upon completion of the sample evaluation and the investigation; a follow-up report will be filed.All information reasonably known as of (b)(6) 2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by halyard health represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to halyard health.Halyard health has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the halyard health complaint database and identified as complaint (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that after 12-days of placement the child was moved in preparation for transport.During the transport of the patient the mic-key device tube snapped.The health professionals tried to replace the mic-key tube but it was impossible because they could not deflate the balloon on the device.The health professionals tried to remove the device three times.On the fourth attempt, they managed to remove 3.9 ml of water.The presence of a white residue which was obstructing the tub was noted.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
One used mickey sample and one syringe filled with 3cc liquid with a bit of white debris floating in it was received.No product packaging was received.The mickey sample appeared in generally clean condition.The mickey was examined and photographed prior to decontamination.No additional white residue was observed.No issues were encountered inflating and deflating the balloon during decontamination and rinse.After decontamination, the sample was examined under magnification.No white residue was found in the inflation port or in the balloon.The balloon was inflated and deflated multiple time without issue.No failure was detected.All information reasonably known as of 04-may-2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by halyard health represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to halyard health.Halyard health has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the halyard health complaint database and identified as complaint (b)(4).
|
| |
|
Search Alerts/Recalls
|