| Model Number |
BC0411XC-ID |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Hemorrhage, Cerebral (1889); Perforation of Vessels (2135)
|
| Date of Event |
01/31/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The lot number was provided.A review of the approved device history records indicated the lot met all release criteria.A lot history trending review was performed and there were no similar complaints for this lot number.The device was not returned for evaluation; therefore, a product analysis could not be performed.The root cause cannot be determined.The ifu identifies vessel perforation and hemorrhage as potential complications associated with use of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the scepter balloon was inflated a first time without any issue and good visibility; however, visibility of the balloon was lost during the 2nd inflation, so control of the pressure was done with a syringe.During this inflation, a rupture of the artery occurred, resulting in an intracerebral hemorrhage.The physician did not observe a device malfunction and does not know the reason for the artery rupture.The patient's outcome after the procedure was reported to be poor.The patient's current status is unknown.
|
| |
|
Search Alerts/Recalls
|