• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. ETHLN BLK 18IN 6-0 S/A P-1 PRM; SUTURE, NONABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. ETHLN BLK 18IN 6-0 S/A P-1 PRM; SUTURE, NONABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number 697H
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Please inform if 1 or 4 samples will be returned for evaluation.What is the lot number of the samples to be returned?.
 
Event or Problem Description
It was reported that a patient underwent an unknown procedure on (b)(6) 2018 and suture was used.It was reported that the suture broke during the procedure.The doctor reported that the suture is very brittle and of very poor quality.No adverse patient outcome was reported.Additional information has been requested.
 
Additional Manufacturer Narrative
Product complaint # pc-(b)(4).The actual sample was returned for evaluation.During the visual inspection of the opened , the swage and attachment area were as expected and marks were observed on the body needle.The end of the suture was observed cut appears to be by use of a surgical instrument.The other section of the suture was not returned.Per condition of the sample, no additional test could be performed.In handling this or any other suture material, care should be taken to avoid damage from handling.Avoid crushing or crimping damage due to application of surgical instruments such as forceps or needle holders.According to the sample condition suggests an improper handling of the sample.A review of the batch manufacturing records was conducted and the batch met all finished goods release criteria.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ETHLN BLK 18IN 6-0 S/A P-1 PRM
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 0151
MDR Report Key7310966
Report Number2210968-2018-71247
Device Sequence Number11077937
Product Code GAR
UDI-Device Identifier10705031008557
UDI-Public10705031008557
Combination Product (Y/N)N
Initial Reporter CountrySA
PMA/510(K) Number
K946173
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 02/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2022
Device Catalogue Number697H
Device Lot NumberLKP660
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2018
Initial Date Received by Manufacturer 02/12/2018
Supplement Date Received by Manufacturer03/16/2018
Initial Report FDA Received Date03/02/2018
Supplement Report FDA Received Date04/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-