| Model Number |
012001 |
| Medical Device Problem Code |
Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/26/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic living donor nephrectomy, the device cut tissue poorly.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The instrument registered a mechanical fault.The instrument was disassembled; a crack was noticed in the proximal portion of the probe shaft.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the cracked probe shaft may occur when the probe shaft makes contact with the outer tube while the system was energized.The information for use for this product states, avoid using the device as a lever, either while dissecting or while activating the instrument.Added stress to the tip may cause the probe to touch in the inner tube portion of the instrument, resulting in mechanical failure and/or rendering the instrument inoperable.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|