• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS 1 MULTIFOCAL; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS 1 MULTIFOCAL; MULTIFOCAL IOLS Back to Search Results
Model Number ZKB00
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: unknown, not provided, but the best estimate date is between (b)(6) 2017.(b)(4).Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a tecnis multifocal intraocular lens (model zkb00 +20.0 diopter) was explanted in a secondary procedure due to the patient experiencing dysphotopsia.A tecnis monofocal intraocular lens (model z9002 +19.5 diopter) was implanted as the replacement lens for the patient.The patient was reported to be doing well at discharge.No further information was provided.
 
Additional Manufacturer Narrative
Additional information: device available for evaluation ¿ yes, returned to manufacturer on 3/15/2018.Device returned to manufacturer ¿ yes.Device evaluation: the intraocular lens (iol) was returned to the manufacturing site for evaluation.The product was received in a lens case.Visual inspection with the unaided eye shows the product is cut in pieces, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.The customer's reported complaint was not verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS 1 MULTIFOCAL
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key7311281
Report Number9614546-2018-00211
Device Sequence Number1250746
Product Code MFK
UDI-Device Identifier05050474560086
UDI-Public(01)05050474560086(17)190126
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date01/26/2019
Device Model NumberZKB00
Device Catalogue NumberZKB00U0200
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2018
Initial Date Received by Manufacturer 02/01/2018
Supplement Date Received by Manufacturer04/30/2018
Initial Report FDA Received Date03/02/2018
Supplement Report FDA Received Date05/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-