Model Number ZCT450
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: unknown, not provided, but the best estimate date is between (b)(6) 2018.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
Initially it was reported that there was a tecnis toric calculator error.The calculation for the intraocular lens came out with +24.5 but the lens should have been +14.5 or thereabouts.Through follow up it was clarified that it was not a calculator error but a data entry error into the calculator by a technician.The implanted lens (model zct450 +24.5 diopter) was subsequently explanted in a secondary procedure.At explant, the incision was enlarged and a model zct300 +15.0 diopter was implanted as a replacement.The patient was reported to be doing good at discharge.No further information was provided.
 
Additional Manufacturer Narrative
Additional information: device available for evaluation ¿ yes, returned to manufacturer on 2/21/2018.Device returned to manufacturer ¿ yes.Device evaluation: the intraocular lens (iol) was returned to the manufacturing site for evaluation.Visual inspection performed by a qualified operator using 12x magnification shows the product is cut in half, most probably to make explant possible, it also showed that a haptic was missing.The lens is contaminated, dust particles are present.This is expected as the lens was transported from controlled environment during manufacturing to a non-sterile, non-environmental controlled area.The customer's reported complaint was not verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
TORIC IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key7311383
Report Number9614546-2018-00212
Device Sequence Number19135707
Product Code HQL
UDI-Device Identifier05050474603073
UDI-Public(01)05050474603073(17)201001
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date10/01/2020
Device Model NumberZCT450
Device Catalogue NumberZCT450U245
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2018
Initial Date Received by Manufacturer 02/01/2018
Supplement Date Received by Manufacturer05/07/2018
Initial Report FDA Received Date03/02/2018
Supplement Report FDA Received Date05/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-