MAUDE Adverse Event Report: STRYKER SPINE-US AVS PL INSERTER 8 AND 9MM; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SPINE-US AVS PL INSERTER 8 AND 9MM; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 48350951
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/08/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that upon starting ratcheting the instrument wasn't disconnecting from the implant and broke.Tight space was observed.The procedure was successfully completed and there are no reports of adverse consequences to the patient.
 
Additional Manufacturer Narrative
Method: visual inspection, device history review, complaint history review, risk assessment; result: upon visual inspection, the tip of the instrument was confirmed to be deformed and the inner shaft was found to be slightly bent.No adverse consequences to the patient was reported.Manufacturing history was reviewed and no issues were identified.The age of the device was found to be more than 13 years.The ifu for instruments states that the instruments may fracture depending on the age of the device, the number of times they are used as well as the precautions taken in handling, cleaning and storage.Conclusion: based on the device inspection and provided information, probable root causes of the reported event are: extensive use of the instrument and normal wear and tear.
 
Event or Problem Description
It was reported that upon starting ratcheting the instrument wasn't disconnecting from the implant and broke.Tight space was observed.The procedure was successfully completed and there are no reports of adverse consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVS PL INSERTER 8 AND 9MM
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER SPINE-US
2 pearl court
allendale NJ 07401
MDR Report Key7311386
Report Number0009617544-2018-00058
Device Sequence Number110504
Product Code LXH
UDI-Device Identifier07613327267914
UDI-Public(01)07613327267914
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number48350951
Device Catalogue Number48350951
Device Lot Number058851
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/12/2018
Initial Date Received by Manufacturer 02/08/2018
Supplement Date Received by Manufacturer05/31/2018
Initial Report FDA Received Date03/02/2018
Supplement Report FDA Received Date06/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-