• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROVENTION, INC. AZUR 18 PERIPHERAL V-TRAK SYSTEM; EMBOLIZATION COIL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION, INC. AZUR 18 PERIPHERAL V-TRAK SYSTEM; EMBOLIZATION COIL Back to Search Results
Model Number 45-480410-L
Medical Device Problem Codes Detachment Of Device Component (1104); Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/30/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The lot number was provided.A review of the device history records indicated the lot met all release criteria.A lot history trending review was performed and there were no similar complaints for this lot number.The device will not be returned for evaluation; therefore, a product analysis will not be performed.The root cause cannot be determined.The submission of this report or related information to the fda, and its release by fda, does not reflect a conclusion by the party submitting this report or the fda that the report or related information is an admission that the manufacturer, their employees, or the device caused or contributed to the reportable event.
 
Event or Problem Description
It was reported that after attempting to position an embolization coil in the arteriovenous malformation, the physician passed a wire down the left anterior descending artery.Resistance was encountered when attempting to advance the coil in the microcatheter.The pusher broke 1cm proximal to the coil attachment point.The detached coil was successfully retrieved with a snare device.There was no reported patient injury.The patient was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AZUR 18 PERIPHERAL V-TRAK SYSTEM
Common Device Name
EMBOLIZATION COIL
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise drive
aliso viejo CA 92656
MDR Report Key7311572
Report Number2032493-2018-00043
Device Sequence Number19071807
Product Code KRD
UDI-Device Identifier00810170017902
UDI-Public(01)00810170017902(11)170127(17)211231(10)170127LC1
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
K091882
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date12/31/2021
Device Model Number45-480410-L
Device Catalogue Number45-480410
Device Lot Number170127LC1
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/02/2018
Initial Report FDA Received Date03/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age53 YR
Patient Weight109
-
-