Model Number 45-750511
Medical Device Problem Codes Detachment Of Device Component (1104); Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The lot number was provided.A review of the approved device history records indicated the lot met all release criteria.A lot history trending review was performed and there were no similar complaints for this lot number.The coil remains implanted in the patient and the remainder of the device was discarded at the user facility; therefore, a product analysis could not be performed.The root cause cannot be determined.The device was used during the same procedure as was reported in mfr.Report# 2032493-2018-00042.
 
Event or Problem Description
It was reported that resistance was encountered while advancing an embolization coil in the microcatheter.During withdrawal, the coil detached in the microcatheter.The coil was pushed into the intended area with a saline injection.There was no reported intervention or patient injury.The patient was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AZUR CX DETACHABLE
Common Device Name
EMBOLIZATION COIL
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise drive
aliso viejo CA 92656
MDR Report Key7311573
Report Number2032493-2018-00041
Device Sequence Number1261080
Product Code KRD
UDI-Device Identifier00812636021027
UDI-Public(01)00812636021027(11)161110(17)211031(10)16111011
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K151358
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date10/31/2021
Device Model Number45-750511
Device Lot Number16111011
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/06/2018
Initial Report FDA Received Date03/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age71 YR
Patient Weight109
-
-