BARD ACCESS SYSTEMS SITE~RITE 8 ULTRASOUND SYSTEM CONSOLE WITH CONNECTOR; SYSTEM, IMAGING, ULTRASONIC
|
Back to Search Results |
|
Model Number 9770600 |
Device Problem
Device Stops Intermittently (1599)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device has not been returned to the manufacturer, at this time, for evaluation.A lot history review (lhr) review is not possible, as the device is manufactured using a unique serial number and not by lot number.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.Device has not yet been returned for evaluation.
|
|
Event Description
|
Per biomed: sr 8 ultrasound system acoustic image intermittently stops and/or lags.After a fresh boot-up the system functionality is normal.After a 10+ minutes of run time the system begins to lag in the image or experience stoppage in the active ultrasound scan.
|
|
Manufacturer Narrative
|
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of lagging ultrasound image was unconfirmed.After initial power up and without adjusting any settings to the unit, no lag was noticed during the evaluation.The root cause of the reported failure is inconclusive as the reported issue could not be reproduced during evaluation.No other functionality issues with the equipment were found during evaluation/servicing.A lot history review (lhr) review is not possible, as the device is manufactured using a unique serial number and not by lot number.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
|
|
Event Description
|
Per biomed: sr 8 ultrasound system acoustic image intermittently stops and/or lags.After a fresh boot-up the system functionality is normal.After a 10+ minutes of run time the system begins to lag in the image or experience stoppage in the active ultrasound scan.
|
|