• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC ELECTRIC PEN DRIVE 60,000 RPM; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESSORIES AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC ELECTRIC PEN DRIVE 60,000 RPM; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESSORIES AND ATTACHMENT Back to Search Results
Catalog Number 05.001.010
Device Problem Device Inoperable (1663)
Patient Problems No Known Impact Or Consequence To Patient (2692); Not Applicable (3189)
Event Date 02/12/2018
Event Type  malfunction  
Manufacturer Narrative
The actual device was returned for evaluation.Reliability engineering evaluated the device and determined that the device had no power; therefore, the pretest could not be completed.Therefore, the reported condition was confirmed.It was further determined that there was corrosion on the motor and the electrical control unit (ecu).The assignable root cause was determined to be due to improper cleaning, which is user error/misuse/abuse.If additional information should become available a supplemental medwatch will be submitted accordingly.
 
Event Description
It was reported that during an unspecified veterinary surgical procedure, it was determined that the electric pen drive device was unresponsive.It was not reported if there was a delay in the surgical procedure, or if a spare device was available for use.There was no human patient involvement as this was a veterinary procedure.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECTRIC PEN DRIVE 60,000 RPM
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESSORIES AND ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES PRODUKTIONS GMBH
hauptstrasse 24
waldenburg 4437
SZ   4437
Manufacturer Contact
michael cote
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key7312317
MDR Text Key101571506
Report Number8030965-2018-51395
Device Sequence Number1
Product Code HWE
UDI-Device Identifier7611819167711
UDI-Public(01)7611819167711(11)070115
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K043310
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 02/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number05.001.010
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2018
Initial Date Manufacturer Received 02/12/2018
Initial Date FDA Received03/04/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/15/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-