DEPUY SYNTHES PRODUCTS LLC ELECTRIC PEN DRIVE 60,000 RPM; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESSORIES AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 05.001.010 |
Device Problem
Device Inoperable (1663)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); Not Applicable (3189)
|
Event Date 02/12/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
The actual device was returned for evaluation.Reliability engineering evaluated the device and determined that the device had no power; therefore, the pretest could not be completed.Therefore, the reported condition was confirmed.It was further determined that there was corrosion on the motor and the electrical control unit (ecu).The assignable root cause was determined to be due to improper cleaning, which is user error/misuse/abuse.If additional information should become available a supplemental medwatch will be submitted accordingly.
|
|
Event Description
|
It was reported that during an unspecified veterinary surgical procedure, it was determined that the electric pen drive device was unresponsive.It was not reported if there was a delay in the surgical procedure, or if a spare device was available for use.There was no human patient involvement as this was a veterinary procedure.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|