| Model Number |
R SERIES |
| Medical Device Problem Code |
Device Operational Issue (2914)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device failed leakage current test.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation and the customer's report was not replicated or confirmed.The device was put through extensive testing without duplicating the report.The device was recertified and returned to the customer.The multi-function cable and ecg cable were not returned as part of this investigation.No trend is associated with reports of this type.
|
| |
|
Search Alerts/Recalls
|