• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HALYARD - IRVINE HYHSURGPN,400X4,P.ONLY,-,OQ,5; ELASTOMERIC LFR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HALYARD - IRVINE HYHSURGPN,400X4,P.ONLY,-,OQ,5; ELASTOMERIC LFR Back to Search Results
Model Number P400X4
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 02/05/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device history record for the reported lot number, 0202775527, in this complaint was reviewed and the material was produced according to the manufacturing procedures and met the quality requirements.One sample device was returned.Upon completion of the sample evaluation and investigation; a follow-up report will be filed.All information reasonably known as of 28-feb-2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by halyard health represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to halyard health.Halyard health has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the halyard health complaint database and identified as (b)(4).
 
Event or Problem Description
Fill volume: 500 ml, flow rate: unknown, procedure: knee surgery, cathplace: knee.It was reported that a catheter break occurred last night.The patient came into office with "stuck catheter," the nurse pulled in different positions and physician's' assistant (pa) pulled in different positions.The catheter snapped off inside the patient's leg right above the last single black mark on eth device.No additional information was provided.
 
Additional Manufacturer Narrative
Additional information: weight.A silver soaker catheter was received attached to a pump.A visual inspection found that catheter was broken at the first black markation and that the infusion segment was not on the catheter, also not returned with the other portion of the catheter.There was evidence of stretching at the 4th marking and continues throughout the catheter to where the breakage occurred.The breakage occurred right before the infusion segment just after the 1st black marking.The outer diameter of the non-stretched area was measured to be 0.039¿ and the stretched area.028¿.The catheter was examined under a microscope magnified at 1x, no signs of brittleness were observed.Although the hollow fiber was broken.Tensile strength was performed on the silver-soaker catheter using the instron machine.The passing rate for the test is >2.8(lbf) at the infusion segment and >4.00 (lbf) for the mid-body segment.The result for the catheter mid-body segment was 9.59(lbf).The infusion segment was not available to test.The catheter portion that was returned met specifications during the tensile test and no additional testing was performed.The investigation summary concluded that silver-soaker catheter was damaged and received not fully intact.Tensile strength was performed on the mid-body segment and met specifications.The excessive force failure mode was chosen because during visual observation of the catheter, stretching and breakage was observed.Tensile strength on the mid-body met specifications.Using excessive force >4.00(lbf) on the catheter at the mid-body segment and >2.8(lbf) at the infusion segment will cause it to break.All information reasonably known as of 06-may-2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by halyard health represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to halyard health.Halyard health has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the halyard health complaint database and identified as complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HYHSURGPN,400X4,P.ONLY,-,OQ,5
Common Device Name
ELASTOMERIC LFR
Manufacturer (Section D)
HALYARD - IRVINE
43 discovery
suite 100
irvine CA 92618
MDR Report Key7313090
Report Number2026095-2018-00022
Device Sequence Number11131678
Product Code MEB
UDI-Device Identifier30680651136757
UDI-Public30680651136757
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K063530
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Nurse
Device Expiration Date01/25/2020
Device Model NumberP400X4
Device Catalogue Number101367503
Device Lot Number0202775527
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 02/06/2018
Supplement Date Received by Manufacturer04/17/2018
Initial Report FDA Received Date03/05/2018
Supplement Report FDA Received Date05/08/2018
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ROPIVICAINE
Patient Age71 YR
Patient Weight102
-
-