• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MEDICAL INC. EXTENSION SET; SET, ADMINISTRATION, INTRA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC. EXTENSION SET; SET, ADMINISTRATION, INTRA Back to Search Results
Catalog Number 471954
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Blood Loss (2597)
Date of Event 02/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The sample has not yet been received and the investigation is ongoing at this time.A follow-up report will be submitted when the results of the investigation become available.
 
Event or Problem Description
While administering the electrolyte solution to a nicu patient, the nurse found that the female luer was cracked causing about 5 ml of blood to leak.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).One (1) used set without packaging was returned for evaluation.Visual examination of the used set noted the green luer adapter to be cracked.No other visual defects were noted.The used set was pressure tested per specification with failing results.Leakage was observed at the crack on the green luer adapter.Although the reported defect was confirmed, the exact root cause could not determined at this time.A review of the discrepancy management system (dsms) database was unable to be performed as the lot number was reported as unknown.While we are unable to confirm the root cause of the reported defect, we will maintain this report for future reference, and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTENSION SET
Common Device Name
SET, ADMINISTRATION, INTRA
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd
allentown PA 18109
MDR Report Key7313185
Report Number2523676-2018-00016
Device Sequence Number1847273
Product Code FPA
UDI-Device Identifier04022495770103
UDI-Public04022495770103
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K942988
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 04/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number471954
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/06/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 02/15/2018
Supplement Date Received by Manufacturer02/15/2018
Initial Report FDA Received Date03/05/2018
Supplement Report FDA Received Date04/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Weight2
-
-