Model Number 5638
Medical Device Problem Code Detachment Of Device Component (1104)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/05/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A manufacturing record review was completed and zero nonconformances were found.The returned product evaluation confirmed that a small piece of the distal tip was separated.The damage found on the returned unit was consistent with over torquing against resistance.
 
Event or Problem Description
Physician was using turnpike to assist in wiring the obtuse marginal vessel.While attempting this multiple times the tip of the turnpike lp broke off inside the vessel.Physician attempted to wire the vessel after this happened so as to retrieve the tip, but w/o success.He was quoted by the staff as saying "he didn't need to retrieve the tip because the vessel was shut down and it wouldn't migrate any further".The patient was alert and oriented after departing the cath lab and no issues since.No patient impact or injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURNPIKE LP
Common Device Name
CATHETER
Manufacturer (Section D)
VASCULAR SOLUTIONS, INC
6464 sycamore court north
minneapolis MN 55369
MDR Report Key7313433
Report Number2134812-2018-00011
Device Sequence Number13903609
Product Code DQY
UDI-Device IdentifierM20656380
UDI-PublicM20656380
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K151981
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date01/04/2020
Device Model Number5638
Device Lot Number618090
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/05/2018
Initial Report FDA Received Date03/05/2018
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
-
-