MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC LCP ONE-THIRD TUBULAR PLATE WITH COLLAR 5 HOLES/57MM; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC LCP ONE-THIRD TUBULAR PLATE WITH COLLAR 5 HOLES/57MM; PLATE, FIXATION, BONE Back to Search Results
Model Number 241.351
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device was used for treatment, not diagnosis.Patient weight not available for reporting.Unknown when condition began.Device is not expected to return for investigation.(b)(4).Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that patient underwent original surgery for treatment of a proximal tibial fracture on (b)(6) 2016.Post-operatively, patient complained that when the weather changes,her feet would feel very cold about the implant site.On (b)(6) 2018, patient underwent hardware removal surgery and the following devices fourteen (14) were removed - one (1) proximal tibia plate, one (1) locking compression tubular plate, one (1) 8 holes locking compression plate, two (2) 4.5 mm cortex screws, four (4) 5.0 mm locking screws and five (5) 3.5 mm cortex screws.There was no reported malfunctions associated with any of the devices.All of the devices were removed easily and the patient is reported to be in stable condition.The hardware removal surgery was completed without any reported delay.This is report 2 of 6 for (b)(4).
 
Additional Manufacturer Narrative
A device history record review was performed for the subject device: part number: 241.351 , lot number: 7205928 , part mfg date: 18-jan-2013 , part exp.Date: n/a , manufacturing location: (b)(4).Dhr record review: a review of the device history record revealed no complaint related anomalies.The device history record shows this lot of lcp one-third tubular plates was processed through the normal manufacturing and inspection operations with no non-conformance noted.This lot met all dimensional and visual criteria at the time of manufacture with no issues documented during the manufacture that would contribute to this complaint condition.The review of the raw material device history record revealed this lot met all specifications with one nonconformance noted.A nonconformance (ncr) is associated with raw material lot 6661221.The ncr was initiated due to the raw material having raw material draw lines.Raw material lot 6661221 was dispositioned as conforms and the raw material was processed.Therefore, ncr is not relevant to this complaint.This raw material lot met all dimensional and visual criteria at the time of manufacture with no issues documented during the manufacture that would contribute to this complaint condition.Raw material lot 6661221 was dispositioned as conforms.Therefore, ncr is not relevant to this complaint.This raw material lot met all dimensional and visual criteria at the time of manufacture.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Device was used for treatment, not diagnosis.Device is a single use device but was not reprocessed or reused.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LCP ONE-THIRD TUBULAR PLATE WITH COLLAR 5 HOLES/57MM
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7313827
Report Number2939274-2018-50871
Device Sequence Number1475465
Product Code HRS
UDI-Device Identifier10886982166852
UDI-Public(01)10886982166852(10)7205928
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K011335
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number241.351
Device Catalogue Number241.351
Device Lot Number7205928
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/05/2018
Supplement Date Received by Manufacturer03/14/2018
04/13/2018
Initial Report FDA Received Date03/05/2018
Supplement Report FDA Received Date04/04/2018
05/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age50 YR
-
-