| Model Number |
8486.621 |
| Medical Device Problem Codes |
Bent (1059); Material Fragmentation (1261)
|
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
02/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Actual device returned to manufacturer and examined on 23feb2018.Visual inspection of forceps revealed they are bent distally and that the jaw part of the forceps is missing.Hardness test was performed on jaw part passed specifications.Most likely broke due excessive force being applied combined with normal wear and tear over time.Device approximately 9 years old.No similar complaints of broken jaw have been received on this device type 8486621.Richard wolf (b)(4) considers this matter closed.However, in the event (b)(4) receives any additional information a follow up report will be submitted to fda.(b)(4).
|
| |
|
Event or Problem Description
|
|
Richard wolf (b)(4) was informed by the user facility that grasping forceps (id #8486.21) broke during use and the jaw part of the device still remains in the patient's carpus.User facility emailed and called to request missing/additional information, no response as of 05mar2018.Manufactured date: 28apr2009, lot size: (b)(4) units, purchase date: 10aug2009.
|
| |
|