Model Number ZXT300
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Capsular Bag Tear (2639); Vitrectomy (2643)
Date of Event 02/06/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If implanted or explanted; give date: n/a (not applicable).The intraocular lens was removed and replaced during the same procedure.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
The surgery center reported that doctor inserted a zxt300 20.0 diopter intraocular lens (iol) when he noticed that the capsular bag was weak and had to take the lens out.It was noted that there was no product quality issue.In follow up, it was reported that there was a capsule tear and a vitrectomy was performed after the lens was used.The lens was removed and a replacement lens from another manufacturer was used during the same surgery.There was no incision enlargement performed and no other medical or surgical intervention was required.Reportedly, there was no patient post-op injury; the patient outcome when discharged was good and met the standard.No additional information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation? yes.Returned to manufacturer on: 3/7/2018.Device returned to manufacturer? yes.Device evaluation: the intraocular lens (iol) was returned to the manufacturer.Visual inspection with the unaided eye shows the product is cut in pieces, most probably to make remove and replacement possible.Considering the condition of the lens additional analysis is not possible.The complaint cannot be verified.Manufacturing record review: manufacturing records were reviewed and the device was manufactured according to specification.A search on complaints history revealed no similar complaints were received for this production order number to date.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY TORIC
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA 92705
MDR Report Key7314986
Report Number9614546-2018-00215
Device Sequence Number1316982
Product Code POE
UDI-Device Identifier05050474606388
UDI-Public(01)05050474606388(17)211121
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date11/21/2021
Device Model NumberZXT300
Device Catalogue NumberZXT300U200
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2018
Initial Date Received by Manufacturer 02/07/2018
Supplement Date Received by Manufacturer04/26/2018
Initial Report FDA Received Date03/05/2018
Supplement Report FDA Received Date05/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
1MTEC30 CARTRIDGE, LOT# CC13806
Outcome Attributed to Adverse Event Required Intervention;
Patient Age75 YR
-
-