Model Number GLJ-53-M
Medical Device Problem Codes Break (1069); Product Quality Problem (1506)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/29/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of the device.The sutures were degraded.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.
 
Event or Problem Description
According to the reporter, during the procedure when the user attempted to use the device, the suture easily broke.The customer also noted that the suture was discolored.The procedure was completed with another device.There was no injury to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYSORB
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key7316214
Report Number1219930-2018-01276
Device Sequence Number16170587
Product Code GAM
UDI-Device Identifier10884521045569
UDI-Public10884521045569
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K963253
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,other,user facility
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 03/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2020
Device Model NumberGLJ-53-M
Device Catalogue NumberGLJ-53-M
Device Lot NumberB5A0444X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/16/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/13/2017
Initial Report FDA Received Date03/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-