| Model Number |
77108 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Hypovolemia (2243)
|
| Date of Event |
01/30/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic is unable to determine the root cause of the issue at the current time without the return of the product.Medtronic is continuing to pursue additional information and has made requests for the return of the device if it is removed from the patient.From other; device has been requested but has not been returned yet.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that during use of a dlp paediatric one-piece arterial cannula approximately 27 minutes after first use of the cannula the anaesthetist noticed the presence of blood coming from a split in the cannula.A cupula was implemented under the canula to recover lost blood.It is unknown if the cannula was replaced.The patient experience hypovolemia.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual analysis: visual inspection shows evidence of a crack in the fitting/connector at the interface between the body and the connector.Performance analysis: during the cleaning process there was a leak observed from the connector/cannula body interface.If information is provided in the future, a supplemental report will be issued.
|
| |
|